Brake Light Switch: Testing and Inspection

2012 Chevrolet Impala V6-3.6LSECTION Testing and Inspection



Brake Pedal Position Sensor Calibration

Calibration Criteria

The brake pedal position sensor (BPPS) receives a low reference signal and a 10-volt reference signal from the body control module (BCM). Whenever the brake pedal is applied, the brake pedal position sensor applies a variable voltage signal to the BCM through the stop lamp switch signal circuit. Brake pedal position sensor calibration must be performed after the brake pedal position sensor or BCM have been serviced. The calibration procedure will set brake pedal position sensor home value. This value is used by the BCM to determine the action of the driver applying the brake system and to provide this information to the vehicle subsystems via the class 2 communication bus.

Calibration Procedure

Important:
If the master brake cylinder has been replaced, verify that the correct pushrod is installed before attempting to calibrate the brake pedal position sensor.


1. Turn ON the ignition, with the engine OFF.

2. Install a scan tool, select the proper vehicle and options.

3. Navigate to the BCM menu.

4. Select Module Setup menu item.

5. Select the BPPS sensor calibration procedure and follow the directions displayed on the screen.

RENDER: 1.0x

NO RELATED

Recommended Tools & Savings

Use the Manual With the Right Hardware

Pair factory procedures with proven DIY tools so the instructions are easier to execute.

Affiliate disclosure: We may earn a commission at no extra cost to you.

When to See a Mechanic

Stop DIY work and contact a certified mechanic immediately if any of the following apply:

  • You smell fuel, burning insulation, or see smoke.
  • Brakes feel soft, pull hard to one side, or make grinding noises.
  • The engine overheats, stalls repeatedly, or misfires under load.
  • You are missing required tools, torque specs, or safe lifting equipment.
  • You are not confident in the next step or safety outcome.